UJI QUALITY CONTROL PESAWAT MAGNETIC RESONANCE IMAGING SIEMENS MAGNETOM CONCERTO 0,2 TESLA DI INSTALASI RADIOLOGI RSUD DR KANUJOSO DJATIWIBOWO BALIKPAPAN
Abstract
Background : Radiology quality control is a part of radiology quality assurance which is directly linked with physical measurements of facility performance and indirectly linked with the expected image quality. The advantage of proceeding with the performance test on MRI device are the consistency of operational testing that integrated of all MRI system and producing high quality image, reducing the possibility of artifact appearance, and also preemptive identification of potential problems.
Methods : The method of this Research is Qualitative Survey. The researchers are doing the survey on Quality Control test on MRI device brand Siemens Magnetom Concerto 0,2T on Radiology department of RSUD. DR. Kanujoso Djatiwibowo Balikpapan.
Results : The whole result of this research shows that MRI device on RSUD. DR. Kanujoso Djatiwibowo Balikpapan has yet to fulfill the international standard.
Conclusion : The whole result of this research shows that MRI device on RSUD. DR. Kanujoso Djatiwibowo Balikpapan has yet to fulfill the international standard on 5 parameter, such as the SNR testing, Artifact Analysis, Geometric Accuracy, High Contrast Resolution and Low Contrast Resolution. On the other hand, 2 other parameter such as slice thickness accuracy and slice position accuracy of the MRI device has already fulfill the standard and have to improve on the 5 parameter which hasn’t fulfill the factory and international standardsKeywords
Full Text:
PDFReferences
AAPM Report No. 100.2010. Acceptance Testingand Quality Assurance Proseduresfor Magnetic Resonance Imanging Facilities. American Association of Physicists in Medicine
American Collage of Radiology (ACR). 2001. Magnetic Resonance Imaging Quality Control Manual. New York.
American Collage of Radiology (ACR). 2009. ACR Technical Standard For Diagnostic Medical Physics Performance Monitoring Of Magnetic Resonance Imaging (MRI)Equipment. New York
American Collage of Radiology (ACR). 2012. NCl Centers of Quantitative Imeging Axcellence . New York
American Collage of Radiology (ACR). 2015. Magnetic Resonance Imaging Quality Control Manual. New York.
Miller, Karl. 2000. Physical Practice in MRI Signal and SNR.
Notosiswoyo, Mulyono. 2004. Pemanfaatan Magnetic Resonance Imaging (MRI) Sebagai Sarana Diagnosa Pasien. Jakarka : Balai Litbang Kesehatan.
Papp, Jeffrey., 2006, Quality Management in The Imaging Sciences, 3th Edition, St. Louis: Mosby Inc
Rasad. Sjahriar, 2005, Radiologi Diagnostik, Cetakan ketujuh. Jakarta : Balai Penerbit FKUI.
Westbrook, Catherine., 2011, Handbook Of MRI Technique Fourth Edition. United Kingdom
Wesbrook, Catherine. 2002. MRI at a Glance. London : Blackwell Scientist Ltd