Analisis Noise Pada Radiografi Thorax Pulmonum Pada Penerapan Modifikasi Faktor Eksposi Aturan 10 kV

Luthfi Rusyadi, Siti Daryati, Dwi Rochmayanti, Andrey Nino Kurniawan

Abstract


Introduction: Quality and dose factors are very important in radiodiagnostics. To produce a constant radiographic quality, the density and contrast produced must remain constant. There is a rule that aims to produce a constant radiographic quality by adding the exposure value, namely the 10 kV rule. This study was conducted to determine the noise in the computed radiography image with the thorax organ produced by modifying the exposure factor of the 10 kV rule and whether it is still within tolerance.

Methods: This quantitative research was conducted with an experimental approach. This is done by taking a series of radiographs that include three exposure factor settings, standard (60 kV, 10 mAs), increased by 10 kV (70 kV, 5 mAs) and lowered by 10 kV (50 kV, 20 mAs). Noise measurement is done by doing ROI in the background area. The exposure index and deviation index values were also recorded as quality and dose references. The data was processed and analyzed by statistical tests.

Results: From the statistical test results, there is a significant relationship between kV and noise with a sig (1-tailed) of ,000. Noise on the standard exposure factor has a lower noise than the modified exposure factor with a difference of 0.2. From the quality aspect, the most optimum exposure index and deviation index indicators are in the range of 70 KV and 5 mAs.

Conclusion: The results of the statistical test of the relationship of kV to noise obtained at 50 Kv and 20 mAs, 60 kV and 10 mAs gave a significance value of 0.263 and 0.435, while at 70 kV and 10 mAs with Sig. (1-tailed) of .000 which means the relationship between kV to noise is strong because the sig value is below 0.05.


Keywords


Citra computed radiography; aturan 10 kV; noise

Full Text:

PDF

References


Arymurthy, A DKK. 1992. Pengantar Pengolahan Citra. PT Elex Media Komputindo Kelompok Gramedia: Jakarta

Ballinger, Philip W. 2003. Merril of Atlas Radiographic Positioning and Radiologic Procedure, Eight Edition Vol. III. Missouri : Mosby, Inc.

Bushong, SC. 2001. Radiologic Science For Technologist Physic, Biology and Protection, Seventh Edition. The CV Mosby Company: Washington DC.

Carlton, Richard R. and Adler M. Arlene. 2001. Principle Radiographic Imaging An Art And Science. JRD Edition. Mosby Company: Toronto.

Mervyn D. Cohen, Matt L. Cooper, Kelly Piersal, Bruce K. Apgar. 2010, Quality assurance : using the exposure index and the deviation index to monitor radiation exposure for postable chest radiograps in neonates, Pediatric Radiology DOI 10.1007/s00247-010-1951-9. © Springer-Verlag 2010

J.A. Seibert, Richard L. Morin. 2011, The standardized exposure index for digital radiography : an opportunity for optimization of radiation dose to the pediatric population. Pediatric Radiology 41:573-581. © Springerlink.com

Papp, Jeffrey.2006. Quality Management In The Imaging Science. Third Edition. Mosby Inc: United States Of America.

Pujiati. 2007. Pengaruh Kenaikan kV aturan 15% Terhadap Kualitas Radiograf dan Paparan Radiasi. DIII Teknik Radiodiagnostik Dan Radioterapi Semarang: Politeknik Kesehatan Kementrian Kesehatan Semarang.

Rasad, S Dkk. 2005. Radiologi Diagnostik. Gaya Baru: Jakarta.

Seibert, dkk. 2006. Acceptance Testing And Quality Control Of Photostimulable Storage Phospor Imaging System. www.aapm.org. Diakses pada tanggal 2 Februari 2016.

Utomo. 2010. Pengaruh Modifikasi Faktor Eksposi dengan Metode Aturan 10 kV Terhadap Kualitas Radiograf (Densitas dan Kontras) Pada Radiograf Ankle Joint. DIII Teknik Radiodiagnostik Dan Radioterapi Semarang: Politeknik Kesehatan Kementrian Kesehatan Semarang.




DOI: https://doi.org/10.31983/jimed.v7i2.7473

Article Metrics

Abstract view : 852
Download PDF : 669

Refbacks

  • There are currently no refbacks.



Creative Commons License  Statcounter Global Stats - Browser, OS, Search Engine including Mobile  Usage Share          Dimensions AI | The most advanced scientific research database

             
JURNAL IMEJING DIAGNOSTIK by http://ejournal.poltekkes-smg.ac.id/ojs/index.php/jimed is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.

View My Stats