PENGARUH VARIASI TIME REPETITION (TR) TERHADAP KUALITAS CITRADAN INFORMASI CITRA PADA PEMERIKSAAN MRI LUMBALSEKUENS T2 FSE POTONGAN SAGITAL

Syafrudin Nizar, Fatimah Fatimah, Irwan Kartili

Abstract


Background : TR (Time Repetition) is an MRI parameter. TR controls the amount of longitudinal magnetization recovery before the next RF pulse. TR parameter is one of parameters that can affect image quality and image information. Long TR arrangements can increase SNR (Signal To Noise Ratio), CNR (Contrast To Noise Ratio) and provide better contrast, but increase the examination time. The purpose of this research is to know the effect of TR to image quality, difference of image information and TR optimal on MRI Lumbal sequence T2 FSE sagittal.

Methods : The type of this research is quantitative research with experimental approach. The sample of research used is 6 probandus. TR variations used are TR 2000, 2500, 3000, 3500, 4000 with other parameters controlled. Aspects of image quality and image information values assessed in this study include Corpus Vertebrae, Discus Vertebrae, Cerebro Spinal Fluid (CSF), Medulla Spinalis, and Soft Tissue. Processing and analysis of image quality data is done by data normality test using shapiro-wilk, then test influence correlation pearson, linear regression and repeated anova. and for image information is done normality data test using shapiro-wilk, and friedman test.

Result : The results of the test analysis showed that there was a TR effect on SNR image quality (p = 0.001) and CNR (p = 0.007) with (p <0.05) on all assessed organs. There is a difference in image information of MRI Lumbal T2 FSE sequence of sagittal pieces on Time Repetition variation with significance value (p <0.05). The optimal TR value on image quality is found on TR 4000 variation with mean of highest mean rank that is 42.82. While the optimal TR value on the image information is on the variation of TR 3000 with the highest mean mean rank is 3.25. TR 3000 is considered to produce better image information and is considered optimal on MRI Lumbal sequences of T2 FSE sagittal fragments compared to other TR variations with a duration of examination time of 2 minutes 30 seconds.

Conclusion: Based on the results that good variation TR used on MRI Lumbal examination sequence T2 FSE sagital view is TR 3000 with best image qualit,  informative image information and short scanning time.


Keywords


Time Repetition (TR); SNR; CNR; Image Information; MRI Lumbal T2 FSE sagital view

Full Text:

PDF

References


Bontrager, Kenneth L. 2014.Textbook of radiographic positioning and related anatomy / Kenneth L. Bontrager, John P. Lampignano ; new photography by James C. Winters. – 8th ed.

Bogduk, Nikolai. 2005 Clinical Anatomy of the Lumbar Spine and Sacrum ; Professor of Pain Medicine, University of Newcastle

Fatimah, J. Dahjono, Metria Riza Sativa. 2015. Optimisasi Field Of View (FOV) Terhadap Kualitas Citra Pada T2WI FSE MRI Lumbal Sagital. Health Polytechnics of Semarang-Indonesia

(MR Imaging of the Spine at 3.0T with T2-Weighted IDEAL Fast Recovery Fast Spin-Echo Technique) Gambar MRI Lumbal Sagital. Diakses tanggal 25 November 2017

Hashemi, H. Ray and Bradley, G. William. 1997. MRI The Basics, Williams & Wilkins Company, USA.

Moeller, Torsten B., M.D. 2003. MRI Parameters and Positioning Georg Thieme Verlag, Rüdigerstraße 14, D-70469 Stuttgart, Germany

Neseth, R., 2000, Procedures and Documentation for CT and MRI, McGraw-Hill, Medical Publishing Division, USA

Notosiswoyo, Mulyono , Susy Suswati. 2004. Pemanfaatan Magnetic Resonance Imaging (MRI) Sebagai Sarana Diagnosa Pasien, Media Litbang Kesehatan Volume XIV Nomor 3

Prastowo, Alan Tanjung, dkk, 2013. Korelasi Nilai Time Repetition (TR) dan Time Echo (TE) terhadap Signal to Noise Ratio (SNR) pada citra MRI, Jurusan Fisika, Universitas Diponegoro

Rossi , Michael, 2014 MR Imaging : Comparison of Ti-Weighted Contrast-En hanced Fat-Suppressed Images with Conventional T2-Weighted and Fast Spin-Echo T2-Weighted Images, Department of Radiology, 5072, Stanford University Medical Center, Stanford, CA.

Sugiyono, Prof.Dr, 2007. Statistika Untuk Penelitian. Cv Alfabeta. Jl, Gegerkalong Hilir Bandung

Supranto, J. 2000. Teknik Sampling untuk Survei dan Eksperimen. Penerbit PT Rineka Cipta, Jakarta.

Snell, R., 1997, Clicnical Anatomy Region, Lippincott Williams & Wilkins

Westbrook, C. and Kaut, C., 2003. MRI In Practise, Second Edition. London: Blackwell Science

Westbrook, Catherine. 2014 Handbook of MRI Technique. Department of Allied Health and Medicine Faculty of Health, Social Care and Education Anglia Ruskin University Cambridge, UK

Westbrook, Catherine. 2016 MRI at Glance Third Edition. Department of Allied Health and Medicine Faculty of Health, Social Care and Education Anglia Ruskin University Cambridge, Us

Woodward, Peggy and Wiliam, W. Arrison, 2001. MRI Optimization. Ahand on approach. McGraw-Hill. Co. USA

Wilmink, J.T, 2010 Lumbar Spinal Imaging in Radicular Pain and Related Conditions © Springer-Verlag Berlin Heidelberg

Weyreuther, Martin. 2007, MRI Atlas Orthopedics and Neurosurgery The Spine, Rontgenabteilung, HELIOS Klinik Emil Behring, Walterhoferstr, 11, 1465 Berlin

Yueniwati, Y. 2014. Prosedur Pemeriksaan Radiologi Untuk Mendeteksi Kelainan dan Cedera Tulang Belakang ; Universitas Brawijaya Press ; Malang




DOI: https://doi.org/10.31983/jimed.v5i2.4473

Article Metrics

Abstract view : 3082
Download PDF : 3403

Refbacks

  • There are currently no refbacks.



Creative Commons License  Statcounter Global Stats - Browser, OS, Search Engine including Mobile  Usage Share          Dimensions AI | The most advanced scientific research database

             
JURNAL IMEJING DIAGNOSTIK by http://ejournal.poltekkes-smg.ac.id/ojs/index.php/jimed is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.

View My Stats