Analisis Informasi Citra Anatomi Antara Penggunaan Aplikasi Constance Level Appearance (CLEAR) dan Non CLEAR Pada Pemeriksaan MRI Cervical Irisan Sagital T2WI FSE Kasus HNP

Dea Febri Mardiyanti, Sugiyanto Sugiyanto, Yeti Kartikasari

Abstract


Background: This study aims to determine the anatomical image information between using CLEAR aplication and without CLEAR (Non CLEAR) aplication on cervical MRI examination with T2WI FSE sagital HNP case and to determine which is the best image between using CLEAR aplication and Non CLEAR aplication on cervical MRI examination with T2WI FSE sagital HNP case.

Methods: This research was a quantitative research with an experimental approach Subjects of this study were eight patients in Siloam Lippo Village Hospital. Images resulted was assessed by radiology physician using questionnairs. Data were analyzed using Wilcoxon test to determine differences of image information and using mean rank test to determine the best image between using CLEAR application and without CLEAR aplication.

Results: This study showed that there were differences anatomical image information between using CLEAR application and without CLEAR aplication on cervical MRI examination with T2WI FSE sagital HNP case, with p = 0,000 (p value <0,05). Cervical MRI examination with CLEAR aplication T2WI FSE produced the best image with the sum of the rank was 300,00 on the negative rank which showed that the image information without CLEAR aplication T2WI FSE lower than  image information with CLEAR aplication T2WI FSE.

Conclusion: There were differences the anatomical image information between using CLEAR application and without CLEAR aplication on cervical MRI examination with T2WI FSE sagital HNP case.


Keywords


cervical MRI examination, CLEAR aplication, without CLEAR aplication, T2WI FSE sagital

Full Text:

PDF


DOI: https://doi.org/10.31983/jimed.v1i2.3160

Article Metrics

Abstract view : 534
Download PDF : 1074

Refbacks

  • There are currently no refbacks.



Creative Commons License  Statcounter Global Stats - Browser, OS, Search Engine including Mobile  Usage Share          Dimensions AI | The most advanced scientific research database

             
JURNAL IMEJING DIAGNOSTIK by http://ejournal.poltekkes-smg.ac.id/ojs/index.php/jimed is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.