Analisis Variasi Slice Thickness Terhadap Informasi Anatomi Potongan Axial Pada Pemeriksaan MSCT Cervical Pada Kasus Trauma

Rizki Aditya Nugroho, Jeffri Ardiyanto, Sigit Wijokongko

Abstract


Background: One step to create a good image is by doing slice thickness reconstruction. Slice thickness is thick slices or pieces of the object checked out. The examination procedure of MSCT in case of Cervical trauma in Radiology Installation RSUD Tugurejo Semarang using slice thickness 3 mm. This contrasts with Seeram (2016), which explained that the examination of adult MSCT Cervical Spine using slice thickness between 1-2 mm. This research aims to know the difference of anatomic information resulted by 5 slice thickness variation in MSCT Cervical Spine and to find out which slice thickness can provide optimum anatomic information.

Methods: The type of this research is quantitative with an experimental approach. Data is performed by reconstructing the 5 images of MSCT cervical of trauma patients using 5 slice thickness variation which is 1 mm, 1.5 mm, 2 mm, 2.5 mm and 3 mm. The final images are evaluated by two Respondents to see the differences in anatomic information and then analyzed by using a different test (Friedman Test) from the SPSS version 24.0.

Results: The results showed there is a  significant difference in anatomic information on the corpus, lamina, spinous process, transverse process and fragment fracture between variations in slice thickness of 1 mm, 1.5 mm, 2 mm, 2.5 mm, and 3 mm on the MSCT Cervical examination with a significance value of 0,000 or p <0.05.

Conclusion: The most optimal slice thickness to provide anatomic information on the cervical MSCT examination is 1 mm slice thickness with a mean rank value of 3.64.


Keywords


MSCT Cervical; Trauma; Slice Thickness

Full Text:

PDF

References


American College of Radiology. 2016. ACR-ASNR-ASSR-SPR Practice Parameter for The performance of Computed Tomography (CT) of The Spine. Revised 2016 (Resolution 15). Akses tanggal 5 januari 2019.

Ardiartana. 2015. Trauma Tulang Leher (Cervical). https://ardiartana.wordpress.com/2015/04/10/trauma-tulang-leher-cervical/ . Akses tanggal 3 Januari 2019.

Ballinger, P. W. 2011. Merill’s Atlas of Radiographic Position and Radiologic Procedures. Twelfth Edition. Volume Three. Mosby: USA.

Bontrager, Kenneth L. 2018. Textbook of Radiographic Positioning and Related Anatomy. Eighth Edition. Elsevier Mosby. St. Louis.

Bushberg, J. T, 2012, The Essential Phisics of Medical Imaging. Third Edition. Lippincot Williams & Wilkins. Philadelphia.

Bushong, C. Stewart. 2013. Radiologic Science For Technologist. Tenth Edition. Elsevier Mosby. St. Louis.

Geyer, L.L., Korner, M., Hempel, R., Deak, Z., Mueck, F.G., Linsenmaier, U., Reiser, M.F., Wirth, S. 2013. Evaluation of a dedicated MDCT protocols using iterative image reconstruction after cervical spine trauma. Akses tanggal 4 Januari 2019.

Habifa. 2017. Pemeriksaan CT Scan Vertebra Cervical Non Kontras. http://panduanctscan64slicesiemens.blogspot.com/2017/02/pemeriksaan-ct-scan-vertebra-cervical_26.html. Akses tanggal 28 November 2018.

Kulama, E. 2004. Scanning Protocols for Multislice CT Scanners. Radiological Sciences Unit (RSU), Hammersmith Hospital, Du Cane Road : London W12 OHS, UK.

Munera, F., Rivas,L.A., Nunez Jr, D.B., Quencer, R.M. 2012. Imaging Evaluation of Adult Spinal Injuries : Emphasis on Multidetector CT in Cervical Spine Trauma. Akses tanggal 4 Januari 2019.

Muttaqin, Arif. 2011. Buku Saku Gangguan Muskoloskeletal Aplikasi Pada Praktik Klinik Keperawatan. Buku kedokteran : EGC. Appleton & Lange.

Pearce, Evelyn C. 2009. Anatomi dan Fisiologi untuk Paramedis. Cetakan Ketigapuluh Tiga. Gramedia : Jakarta.

Raman, S.P., Mahesh, M., Blasko, R.V., Fishman, E.K. 2013. CT Scan Parameters and Radiation Dose: Practical Advice for Radiologists. Akses tanggal 6 Januari 2019.

Seeram, Euclid. 2016. Computed Tomography : Physical Principles, Clinical Applications, and Quality Control. Fourth Edition. Elsevier Mosby. St. Louis.




DOI: https://doi.org/10.31983/jimed.v6i2.5824

Article Metrics

Abstract view : 1474
Download PDF : 1481

Refbacks

  • There are currently no refbacks.



Creative Commons License  Statcounter Global Stats - Browser, OS, Search Engine including Mobile  Usage Share          Dimensions AI | The most advanced scientific research database

             
JURNAL IMEJING DIAGNOSTIK by http://ejournal.poltekkes-smg.ac.id/ojs/index.php/jimed is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.

View My Stats