Analisa Informasi Citra MRI Cervical Potongan Sagital Pada Variasi Nilai Time Repetition (TR) Dengan Sekuens Short Tau Inversion Recovery (STIR)

Siti Masrochah, Fatimah Fatimah, Nurdianty Yunitaningrum

Abstract


Background: STIR sequences used in the cervical spine due to the cervical spine is surrounded by lots of fluids and fat, to reinforce the picture of the cervical spine. According to Braun in 2003, TR on STIR used is 4000 ms. Meanwhile, according to Erika J. Ulbribh in 2011, TR on STIR sequences was used that 4860 ms. Based on the writer's observation in several hospitals with the same flight characteristics, the value of TR on STIR sequences different - different. At one hospital using TR 5000 ms, then at other hospitals using the TR 3000 ms. TR value is used based on the theory of 2000 ms.

Methods: This research is a quantitative study with an experimental approach. Data were collected from May to June 2016 in the. The subjects of this study are volunteers by using a variation of the value of Time Repetition. It is caused during data capture none of the patients with the diagnosis of the lesion or tumour. image of five variations Repetition Time value is 4090 ms, 5190 ms, 6290 ms, 7390 ms and 8490 ms. Selection of variation based on the value of TR used.

Results: From the statistical test result known that There are differences in image information MRI Cervical on the variation of the Time Repetition 4090 ms, TR 5190 ms, TR 6290 ms, TR 7390 ms, and TR 8490 ms to the value of ρ value of 0.000 (ρ <0, 05).

That is because Repetition Time (TR) sequences Inversion Recovery (IR) should be long enough to provide opportunities so that Net Magnetization Vectors (NMV) can be in full recovery before the next 180 RF pulse inversion. If TR is too short then each network will be recovery with different levels which will ultimately affect the weighting (weighting) is generated.

Conclusion: There are differences in Cervical MRI image information on the variation of the value of 4090 ms Repetition Time, TR 5190 ms, TR 6290 ms, TR 7390 ms, and TR 8490 ms to the value of ρ value of 0.000 (ρ <0.05).


Keywords


Cervical MRI; Repetition Time (TR); STIR; Image Information

Full Text:

PDF

References


Braun, J, X. Baraliakos, W. Golder, J. Brandt, M. Rudwaleit, J. Listing, M. Bollow, J. Sieper, dan der Heide, 2003. Magnetic Resonance Imaging Exaination of the Spine in Patiens With Anklylosing Spondylitis, Before adn After Successful Therapy With Infliximab, American College of Rheumatology.

Edelman, RR, 2006. Magnetic Tesonance TIP-MRI Database Cervical Spine MRI. Magnetic Resonance-Technology Information Portal ; mri-tip.com.

Hashemi, H. Ray and Bradley, G. William. 1997. MRI : The Basic. Williams & Wilkins, USA.

J. Applegate, E, 1991. Sectional Anatomy Learning Syste, W.B. saunders Company

J Blink, E, 2004. Basic MRI Physics.

Mc. Robbie, Donald W, Elizbeth A. Moore, Martin J. Graves, Martin R. Prince, 2003. MRI From Picture To Proton, Second Edition, Cambridge University Press.

Meidl, T, S. Weckbah, O. Dietrich, M. Reiser dan S.O. Schoenberg, 2009. Magnetic Resonance Imaging of The Cervical Spine: Comparison of 2D T2-Weighted Turbo Spin Echo, 2D T2* Weighted Gradient-Recalled Echo And 3D T2-Weighted Variable Flip-Angle Turbo Spin Echo Sequences, Europan Society Radiology.

Moeler, M.D, torsten B; Reif, M.D, Emil, 2003. MRI Parameter and Positioning, Thieme, Stutgart, New York.

Ness Alver, 1997, all you reallyneed to knpw about MRI physics, University of Maryland Medical Center, USA.

Sim, Julius dan Chris C. Wraight, 2005. The Kappa Statistic in Realibity Studies : Use, Interpretation, and Sample Size Requirements, Physical Therapghy volume 85.

Snell, Rihard S. 1997. Anatomi Klinik Untuk Mahasiswa Kedokteran, Bagian Tiga, Edisi Tiga, Akih Bahasa ; dr. Jan Tambayong. EGC, Jakarta.

Strak, David D, M.D. and Bradley William G, 1998, Magnetic resonance imaging, Mosby, Westline Industrial Drive, St. Louis, Missouri.

Sopiyudin Dahlan, M, 2013. Statistik Untuk Kedokteran dan Kesehatan, Edisi 5, Jakarta.

Ulbrich, Erika J, MD, Suzanne E. Anderson, MD, Andre Busato, PhD, Susanne Abderhalden, MD, Chris Boescch, MD PhD, Heinz Zimmermann, MD, Paul Heini, MD, Juerg Hodler, MD, PhD, dan Matthias Sturzenegger, MD, 2011. Cervical Muscle Area Measurements in Acute Whiplash Patients and Controls, Journal of Magnetic Resonance Imaging.

Wesbrook, C, 1999. Handbook of MRI technique, Blacwell Science Ltd., United Kingdom.

Westbrook, C, and Kaunt, C, 1998.MRI In Pratise, Blackwell Sciene Ltd, United Kingdom.

Woordward, Peggy and William W. Orrison, 1997. A hands-on Approach. Mc. Graw Hill : New York.




DOI: https://doi.org/10.31983/jimed.v6i2.5822

Article Metrics

Abstract view : 700
Download PDF : 520

Refbacks

  • There are currently no refbacks.



Creative Commons License  Statcounter Global Stats - Browser, OS, Search Engine including Mobile  Usage Share          Dimensions AI | The most advanced scientific research database

             
JURNAL IMEJING DIAGNOSTIK by http://ejournal.poltekkes-smg.ac.id/ojs/index.php/jimed is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.

View My Stats